Accuracy and precision of direct bone conduction measurements

1.

Department of Audiology, The Sahlgrenska Academy, Institute of Neuroscience and Physiology, Göteborg, Sweden

2.

Department of Audiology, Hearing activities organization, Habilitation & Health, Göteborg, Sweden

3.

Department of Research and Applications, Cochlear Bone Anchored Solutions AB, Mölnlycke, Sweden

4.

Department of Audiology, Freeman Hospital, Newcastle, United Kingdom

5.

Hörzentrum Oldenburg GmbH, Oldenburg, Germany

B-ENT 2016; 12: 41-51
Read: 994 Downloads: 752 Published: 03 February 2020

Accuracy and precision of direct bone conduction measurements. Objectives: Variability in Baha® sound processor fittings arise from several sources including the nature of the implant-to-bone transmission and transcranial attenuation in patients with single-sided sensorineural deafness. One method of improving the predictability of Baha fittings is to measure the individual patient’s actual bone conduction thresholds via the implant, thereby removing the influence of skin thickness and/or implant location site.

Methodology: One hundred thirty eight adult wearers of the Baha System participated in the study. Direct bone conduction thresholds were obtained through the BC Direct feature of the Cochlear™ Baha Fitting Software combined with the CochlearBaha BP100 sound processor. Test-retest reliability measurement was performed in 58 participants.

Results: Improved transmission of sound through the implant rather than transcutaneously through the skin was confirmed. On average, the BC Direct thresholds were closer to the patient’s unmasked thresholds than the masked values. In patients with single-sided sensorineural deafness, BC Direct results were elevated compared to the contralateral bone conduction thresholds due to transcranial attenuation. The test-retest reliability for the BC Direct measurements was within ±5 dB, which is in within the accepted variability for audiometric test measurements.

Conclusions: Direct bone conduction measurement provides a validated method of comparing the transcutaneous thresholds as measured through audiometry with the percutaneous responses from the Baha sound processor. The Baha fitting based on direct measurements of bone conduction may require less fine-tuning and provide a greater understanding of the variability of the bone conduction sound pathway.